<?xml version="1.0" encoding="UTF-8"?>
<article article-type="research-article" dtd-version="1.3" xml:lang="ru" xmlns:xlink="http://www.w3.org/1999/xlink" xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xsi="http://www.w3.org/2001/XMLSchema-instance" xsi:noNamespaceSchemaLocation="https://metafora.rcsi.science/xsd_files/journal3.xsd">
  <front>
    <journal-meta>
      <journal-id journal-id-type="publisher-id">moitvivt</journal-id>
      <journal-title-group>
        <journal-title xml:lang="ru">Моделирование, оптимизация и информационные технологии</journal-title>
        <trans-title-group xml:lang="en">
          <trans-title>Modeling, Optimization and Information Technology</trans-title>
        </trans-title-group>
      </journal-title-group>
      <issn pub-type="epub">2310-6018</issn>
      <publisher>
        <publisher-name>Издательство</publisher-name>
      </publisher>
    </journal-meta>
    <article-meta>
      <article-id pub-id-type="doi">10.26102/2310-6018/2026.62.11.016</article-id>
      <article-id pub-id-type="custom" custom-type="elpub">2645</article-id>
      <title-group>
        <article-title xml:lang="ru">Раннее выявление нарушений двигательной активности плода: новое носимое устройство на основе пассивных датчиков</article-title>
        <trans-title-group xml:lang="en">
          <trans-title>Early detection of fetal motor activity disorders: a novel wearable device based on passive sensors</trans-title>
        </trans-title-group>
      </title-group>
      <contrib-group>
        <contrib contrib-type="author">
          <contrib-id contrib-id-type="orcid">0009-0005-2722-5976</contrib-id>
          <name-alternatives>
            <name name-style="eastern" xml:lang="ru">
              <surname>Малышев</surname>
              <given-names>Никита Андреевич</given-names>
            </name>
            <name name-style="western" xml:lang="en">
              <surname>Malyshev</surname>
              <given-names>Nikita Andreevich</given-names>
            </name>
          </name-alternatives>
          <email>malyshev.nikita.2016@gmail.com</email>
          <xref ref-type="aff">aff-1</xref>
        </contrib>
      </contrib-group>
      <aff-alternatives id="aff-1">
        <aff xml:lang="ru">Петрозаводский государственный университет</aff>
        <aff xml:lang="en">Petrozavodsk State University</aff>
      </aff-alternatives>
      <pub-date pub-type="epub">
        <day>01</day>
        <month>01</month>
        <year>2026</year>
      </pub-date>
      <volume>1</volume>
      <issue>1</issue>
      <elocation-id>10.26102/2310-6018/2026.62.11.016</elocation-id>
      <permissions>
        <copyright-statement>Copyright © Авторы, 2026</copyright-statement>
        <copyright-year>2026</copyright-year>
        <license license-type="creative-commons-attribution" xlink:href="https://creativecommons.org/licenses/by/4.0/">
          <license-p>This work is licensed under a Creative Commons Attribution 4.0 International License</license-p>
        </license>
      </permissions>
      <self-uri xlink:href="https://moitvivt.ru/ru/journal/article?id=2645"/>
      <abstract xml:lang="ru">
        <p>Непрерывный мониторинг двигательной активности плода является важным инструментом оценки его внутриутробного состояния. Традиционное ультразвуковое наблюдение не подходит для длительного домашнего использования, а субъективный метод материнского восприятия обладает низкой точностью. Акселерометрические системы рассматриваются как альтернатива, однако оптимальное количество датчиков и методы обработки сигналов остаются предметом исследований. Целью работы является разработать и валидировать носимое 4-канальное акселерометрическое устройство для регистрации шевелений плода на сроках гестации 26–40 недель. Устройство содержит четыре датчика LIS331DLH на базе микроконтроллера Raspberry Pi 4. Данные обрабатываются в среде SciLab. Проведено сравнительное исследование с участием 10 беременных женщин (26–40 недель). Референтные методы – ультразвуковое исследование (УЗИ) и субъективная регистрация матерью. Длительность каждой сессии составила 20 минут. По отношению к УЗИ (эталонный метод) совпадение детекций для субъективной материнской оценки составило 63,08 %, для разработанной акселерометрической системы – 92,31 %. Система показала высокую эффективность на всех сроках гестации, особенно, на поздних (40 недель). Предлагаемое устройство по точности значимо превосходит субъективный метод материнской оценки и сопоставимо с ультразвуковым исследованием, что делает его перспективным для домашнего скринингового мониторинга.</p>
      </abstract>
      <trans-abstract xml:lang="en">
        <p>Continuous monitoring of fetal motor activity is an important tool for assessing intrauterine condition. Traditional ultrasound observation is unsuitable for long-term home use, while the subjective method of maternal perception has low accuracy. Accelerometer-based systems are considered an alternative; however, the optimal number of sensors and signal processing methods remain subjects of research. The aim of this work is to develop and validate a wearable 4-channel accelerometric device for recording fetal movements at gestational ages of 26–40 weeks. The device comprises four LIS331DLH sensors based on a Raspberry Pi 4 microcontroller. Data are processed in the SciLab environment. A comparative study was conducted involving 10 pregnant women (26–40 weeks). Reference methods were ultrasound examination (US) and subjective maternal registration. Each session lasted 20 minutes. Relative to ultrasound (the gold standard), the detection agreement for subjective maternal assessment was 63.08 %, while for the developed accelerometric system it was 92.31 %. The system demonstrated high efficacy across all gestational ages, particularly at late stages (40 weeks). The proposed device significantly outperforms the subjective method of maternal assessment in accuracy and is comparable to ultrasound examination, making it promising for home-based screening monitoring.</p>
      </trans-abstract>
      <kwd-group xml:lang="ru">
        <kwd>фетальный мониторинг</kwd>
        <kwd>двигательная активность плода</kwd>
        <kwd>функциональное состояние плода</kwd>
        <kwd>пренатальная диагностика</kwd>
        <kwd>носимое устройство</kwd>
        <kwd>акселерометры</kwd>
        <kwd>регистрация сигналов</kwd>
      </kwd-group>
      <kwd-group xml:lang="en">
        <kwd>fetal monitoring</kwd>
        <kwd>fetal motor activity</kwd>
        <kwd>fetal functional state</kwd>
        <kwd>prenatal diagnosis</kwd>
        <kwd>wearable device</kwd>
        <kwd>accelerometers</kwd>
        <kwd>signal recording</kwd>
      </kwd-group>
      <funding-group>
        <funding-statement xml:lang="ru">Исследование выполнено без спонсорской поддержки.</funding-statement>
        <funding-statement xml:lang="en">The study was performed without external funding.</funding-statement>
      </funding-group>
    </article-meta>
  </front>
  <back>
    <ref-list>
      <title>References</title>
      <ref id="cit1">
        <label>1</label>
        <mixed-citation xml:lang="ru">Репина Е.С.,  Костелей Я.В., Буреев А.Ш., Жданов Д.С., Юрьев С.Ю. Современные возможности дистанционного мониторирования состояния плода. Мать и дитя в Кузбассе. 2022;(4):12–17. https://doi.org/10.24412/2686-7338-2022-4-12-17</mixed-citation>
      </ref>
      <ref id="cit2">
        <label>2</label>
        <mixed-citation xml:lang="ru">Lai J., Nowlan N.C., Vaidyanathan R., Shaw C.J., Lees C. Fetal movements as a predictor of health. Acta obstetricia et Gynecologica Scandinavica. 2016;95(9):968–975. https://doi.org/10.1111/aogs.12944</mixed-citation>
      </ref>
      <ref id="cit3">
        <label>3</label>
        <mixed-citation xml:lang="ru">Scher M.S., Eyre H., Donn S., et al. An interdisciplinary fetal neonatal neurology collaborative promotes integrative life-course brain health. Frontiers in Neurology. 2026;16:1725289. https://doi.org/10.3389/fneur.2025.1725289</mixed-citation>
      </ref>
      <ref id="cit4">
        <label>4</label>
        <mixed-citation xml:lang="ru">Репина Е.С., Костелей Я.В., Буреев А.Ш. и др. Мониторирование внутриутробного состояния плода. История вопроса. Новые возможности фонокардиографии. Бюллетень сибирской медицины. 2023;22(3):141–149. https://doi.org/10.20538/1682-0363-2023-3-141-149</mixed-citation>
      </ref>
      <ref id="cit5">
        <label>5</label>
        <mixed-citation xml:lang="ru">De Vries J.I.P., Visser G.H.A., Prechtl H.F.R. The emergence of fetal behaviour. I. Qualitative aspects. Early Human Development. 1982;7(4):301–322. https://doi.org/10.1016/0378-3782(82)90033-0</mixed-citation>
      </ref>
      <ref id="cit6">
        <label>6</label>
        <mixed-citation xml:lang="ru">Winje B.A., Saastad E., Gunnes N., et al. Analysis of ‘count‐to‐ten’ fetal movement charts: a prospective cohort study. BJOG: An International Journal of Obstetrics &amp; Gynaecology. 2011;118(10):1229–1238. https://doi.org/10.1111/j.1471-0528.2011.02993.x</mixed-citation>
      </ref>
      <ref id="cit7">
        <label>7</label>
        <mixed-citation xml:lang="ru">Khlif M.S.H., Boashash B., Layeghy S., Ben-Jabeur T., Colditz P.B., East Ch. A passive DSP approach to fetal movement detection for monitoring fetal health. In: 2012 11th International Conference on Information Science, Signal Processing and their Applications (ISSPA), 02–05 July 2012, Montreal, QC, Canada. IEEE; 2012. P. 71–76. https://doi.org/10.1109/ISSPA.2012.6310647</mixed-citation>
      </ref>
      <ref id="cit8">
        <label>8</label>
        <mixed-citation xml:lang="ru">Layeghy S., Azemi Gh., Colditz P., Boashash B. Classification of fetal movement accelerometry through time-frequency features. In: 2024 8th International Conference on Signal Processing and Communication Systems (ICSPCS), 15–17 December 2014, Gold Coast, QLD, Australia. IEEE; 2014. https://doi.org/10.1109/ICSPCS.2014.7021055</mixed-citation>
      </ref>
      <ref id="cit9">
        <label>9</label>
        <mixed-citation xml:lang="ru">Ghosh A.K., Catelli D.S., Wilson S., Nowlan N.C., Vaidyanathan R. Multi-modal detection of Fetal Movements Using a Wearable Monitor. Information Fusion. 2024;103:102124. https://doi.org/10.1016/j.inffus.2023.102124</mixed-citation>
      </ref>
      <ref id="cit10">
        <label>10</label>
        <mixed-citation xml:lang="ru">Mesbah M., Khlif M.S., Layeghy S., et al. Automatic fetal movement recognition from multi-channel accelerometry data. Computer Methods and Programs in Biomedicine. 2021;210:106377. https://doi.org/10.1016/j.cmpb.2021.106377</mixed-citation>
      </ref>
      <ref id="cit11">
        <label>11</label>
        <mixed-citation xml:lang="ru">Сорокин Г.А. Фильтры нижних частот. Вестник Южно-Уральского государственного университета. Серия: Компьютерные технологии, управление, радиоэлектроника. 2015;15(1):100–107.</mixed-citation>
      </ref>
      <ref id="cit12">
        <label>12</label>
        <mixed-citation xml:lang="ru">Boashash B., Khlif M.S., Ben-Jabeur T., East C.E., Colditz P.B. Passive detection of accelerometer-recorded fetal movements using a time-frequency signal processing approach. Digital Signal Processing. 2014;25:134–155. https://doi.org/10.1016/j.dsp.2013.10.002</mixed-citation>
      </ref>
    </ref-list>
    <fn-group>
      <fn fn-type="conflict">
        <p>The authors declare that there are no conflicts of interest present.</p>
      </fn>
    </fn-group>
  </back>